首页 / 院系成果 / 成果详情页

Thirty-day Outcomes of First-in-man Implantation of a Novel Transcatheter Edge-to-edge Mitral Repair System in Patients With Severe Mitral Regurgitation  期刊论文  

  • 编号:
    3762FC53EED9B1470FB117A977AC46B4
  • 作者:
    Xu, Kai#[1]Wang, Yan#[2]Chen, Shaoliang#[3]Pan, Xiangbin#[4]He, Ben[5];Zhang, Ruiyan[6];Wang, Bin[1,2];Zhang, Junjie[3];Zhu, Zhengbin[6];Wang, Bin[1,2];Zhu, Da[4];Han, Yaling*[1]
  • 语种:
    英文
  • 期刊:
    CARDIOLOGY DISCOVERY ISSN:2096-952X 2024 年 4 卷 2 期 (142 - 147) ; JUN
  • 收录:
  • 关键词:
  • 摘要:

    Objectives: The aim of this multicenter, prospective, single-arm pilot study (ClinicalTrials.gov number: NCT05040074) was to observe the procedural and 30-day results of the novel transcatheter mitral valve repair system, SQ-Kyrin (R)-M Clip (Shenqi Medical, Shanghai, China), in patients with severe mitral regurgitation (MR). Methods: The heart team considered patients from 5 centers in China with clinically significant functional mitral regurgitation >= 3+ despite optimal medical therapy or degenerative mitral regurgitation >= 3+ with high surgical risk as candidates for transcatheter repair. All patients received transcatheter edge-to-edge mitral valve repair under general anesthesia. The primary outcome was technical success, which included all of the following measured at the exit from the catheterization laboratory: (1) absence of procedural mortality; (2) successful access, delivery, and retrieval of the device delivery system; (3) successful deployment and correct positioning of the first intended device; and (4) no emergency surgery or reintervention related to the device or access procedure. The secondary outcomes included all-cause mortality, serious adverse events, device success, and procedural success 30 d after the intervention. Results: From June 2021 to December 2021, 18 patients were enrolled in this study with age (75.7 +/- 7.4) years. Fifteen patients had MR 4+, while 3 had MR 3+. Technical success was achieved in all patients, including 6 degenerative mitral regurgitation and 12 functional mitral regurgitation patients. There was no all-cause mortality at 30 d. One patient had single leaflet device attachment within 30 d, which was regarded as a serious adverse event, and the patient was successfully treated with reintervention by implanting another clip. Another patient''s transmitral gradient was 6 mmHg (>5 mmHg), with an effective orifice area of 2.57 cm2 after the procedure. Sixteen patients had device success and procedural success at 30 d postoperation. Fourteen patients had MR 1+, 3 had MR 2+, and only 1 patient had MR 3 + 30 d after the procedure. Conclusions: The results of this feasibility study showed the efficacy and safety of the SQ-Kyrin (R)-M device in the Chinese population with severe MR, laying a solid foundation for a subsequent large-scale confirmatory study.

  • 推荐引用方式
    GB/T 7714:
    Xu Kai,Wang Yan,Chen Shaoliang, et al. Thirty-day Outcomes of First-in-man Implantation of a Novel Transcatheter Edge-to-edge Mitral Repair System in Patients With Severe Mitral Regurgitation [J].CARDIOLOGY DISCOVERY,2024,4(2):142-147.
  • APA:
    Xu Kai,Wang Yan,Chen Shaoliang,Pan Xiangbin,&Han Yaling.(2024).Thirty-day Outcomes of First-in-man Implantation of a Novel Transcatheter Edge-to-edge Mitral Repair System in Patients With Severe Mitral Regurgitation .CARDIOLOGY DISCOVERY,4(2):142-147.
  • MLA:
    Xu Kai, et al. "Thirty-day Outcomes of First-in-man Implantation of a Novel Transcatheter Edge-to-edge Mitral Repair System in Patients With Severe Mitral Regurgitation" .CARDIOLOGY DISCOVERY 4,2(2024):142-147.
  • 入库时间:
    2024/12/2 21:42:08
  • 更新时间:
    2024/12/23 22:30:07
  • 条目包含文件:
    文件类型:PDF,文件大小:
    正在加载全文
浏览次数:11 下载次数:0
浏览次数:11
下载次数:0
打印次数:0
浏览器支持: Google Chrome   火狐   360浏览器极速模式(8.0+极速模式) 